L/K-Einfangverhältnis von65Zn |
| |
Authors: | D Totzek K -W Hoffmann |
| |
Institution: | 1. Institut für Strahlenphysik der Technischen Hochschule Stuttgart, Deutschland
|
| |
Abstract: | A multiwire proportional counter was constructed to measureL/K-capture ratios. The length of the counter is 90 cm, the diameter 14 cm, and it works with pressures from 1 to 6 atm. — The counter was proved by measuring theL/K-capture ratio of37Ar. The result obtained is 0,098±0,003, in agreement with the calculated ratio 0,100 ofBahcall and the results of many other measurements with multiwire proportional counters. — Gaseous radioactiv Zincdiethyl was the substance to determine theL/K-capture ratio of65Zn. The result obtained is 0,111±0,006, in agreement with the calculated ratio 0,109 ofBahcall, while 0,119±0,007 measured with a multiwire proportional counter bySantos Ocampo andConway is somewhat above the calculated ratio. |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |
|