Multi-channel metabolic imaging, with SENSE reconstruction, of hyperpolarized [1-(13)C] pyruvate in a live rat at 3.0 tesla on a clinical MR scanner
Authors:Tropp James  Lupo Janine M  Chen Albert  Calderon Paul  McCune Don  Grafendorfer Thomas  Ozturk-Isik Esin  Larson Peder E Z  Hu Simon  Yen Yi-Fen  Robb Fraser  Bok Robert  Schulte Rolf  Xu Duan  Hurd Ralph  Vigneron Daniel  Nelson Sarah
Institution:GE Healthcare Technologies, Global Applied Sciences Lab, Fremont, CA, United States. james.tropp@med.ge.com
Abstract:We report metabolic images of (13)C, following injection of a bolus of hyperpolarized 1-(13)C] pyruvate in a live rat. The data were acquired on a clinical scanner, using custom coils for volume transmission and array reception. Proton blocking of all carbon resonators enabled proton anatomic imaging with the system body coil, to allow for registration of anatomic and metabolic images, for which good correlation was achieved, with some anatomic features (kidney and heart) clearly visible in a carbon image, without reference to the corresponding proton image. Parallel imaging with sensitivity encoding was used to increase the spatial resolution in the SI direction of the rat. The signal to noise ratio in was in some instances unexpectedly high in the parallel images; variability of the polarization among different trials, plus partial volume effects, are noted as a possible cause of this.