首页 | 本学科首页   官方微博 | 高级检索  
     检索      


ICP-OES determination of traces of Ru, Au, V and Ti preconcentrated and separated by a new poly(epoxy-melamine) chelating resin from solutions
Authors:Xi-Jung Chang  Bo-Lin Gong  Zhi-Xing Su  Dong Yang and Xing-Yin Luo
Institution:(1) Department of Chemistry, Lanzhou University, Lanzhou 730000, P. R. China, CN;(2) Department of Chemistry, Ningxia University, Yinchuan 750021, P. R. China, CN
Abstract:A new poly(epoxy-melamine) chelating resin is synthesized from epoxy resin and used for the preconcentration and separation of traces of Ru(III), Au(III), V(V) and Ti(IV) ions from sample solutions. The ions analyzed can be quantitatively enriched by the resin at a flow-rate of 2 mL/min at pH 4, and quantitatively desorbed with 10 mL of 1 mol/L HCl + 0.2 g CS(NH2)2 at a flow-rate of 1 mL/min with recoveries of over 97%. The chelating resin can be reused 7 times without obvious loss of efficiency. Thousand-fold excesses of coexistent ions caused little interference during the enrichment and determination steps. The RSDs for the determination of 50 ng/mL Ru(III) and Au(III), 5.0 ng/mL V(V) and Ti(IV) were in the range of 1.5–4.5%. The recoveries of added standards in a real sample solution are between 96% and 100%, and the results for the ions analyzed in a nickel alloy sample are in good agreement with their reported values. Received: 12 May 1997 / Revised: 1 September 1997 / Accepted: 9 October 1997
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏