首页 | 本学科首页   官方微博 | 高级检索  
     


Ion-exchange method for separation and concentration of platinum and palladium for analysis of environmental samples by inductively coupled plasma atomic emission spectrometry
Authors:Petya Kovacheva  Rumyana Djingova
Affiliation:Faculty of Chemistry, University of Sofia, 1 James Bourchier Boulevard, 1164 Sofia, Bulgaria
Abstract:An ion-exchange procedure is proposed for determination of Pt and Pd in environmental samples, using a Dowex 1-X10 anion-exchange resin. Pt and Pd were separated from the matrix elements in the sample by selective retention on the column as anionic chloro complexes and subsequent elution by circulated thiourea at 60 °C. The eluent, containing Pt and Pd was analyzed by inductively coupled plasma (ICP) atomic emission spectrometry (AES). Average recoveries of 98% and detection limit of 15 ng/g for both metals were achieved. Analysis of Pt and Pd concentrations in road dust, sampled from several sites in Germany was performed. The comparison of the obtained data with the concentrations of Pt and Pd in the same samples, determined by ICP-MS showed a very good agreement.
Keywords:Platinum   Palladium   Anion-exchange   Environmental samples   ICP-AES
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏