首页 | 本学科首页   官方微博 | 高级检索  
     


Development and validation of a reversed-phase ion-pair high-performance liquid chromatographic method for the determination of risedronate in pharmaceutical preparations
Authors:Kyriakides Demetra  Panderi Irene
Affiliation:University of Athens, School of Pharmacy, Division of Pharmaceutical Chemistry, Panepistimiopolis, Zografou 157 71, Athens, Greece
Abstract:
A stability indicating, reversed-phase ion-pair high-performance liquid chromatographic method was developed and validated for the determination of risedronate in pharmaceutical dosage forms. The determination was performed on a BDS C18 analytical column (250 mm × 4.6 mm i.d., 5 μm particle size); the mobile phase consisted of 0.005 M tetrabutylammonium hydroxide and 0.005 M pyrophosphate sodium (pH 7.0) mixed with acetonitrile in a ratio (78:22, v/v) and pumped at a flow rate 1.00 mL min−1. The ultraviolet (UV) detector was operated at 262 nm. The retention times of magnesium ascorbyl phosphate, which was used as internal standard and risedronate were 4.94 and 5.95 min, respectively. The calibration graph was ranged from 2.50 to 20.00 μg mL−1, while detection and quantitation limits were found to be 0.48 and 1.61 μg mL−1, respectively. The intra- and inter-day percentage relative standard deviations, %R.S.D., were less than 5.9%, while the relative percentage error, %Er, was less than 0.4%. The method was applied to the quality control of commercial tablets and content uniformity test and proved to be suitable for rapid and reliable quality control.
Keywords:Risedronate   Magnesium ascorbyl phosphate   Ion-pair liquid chromatography   Pharmaceutical dosage forms   Stability indicating
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号