Simultaneous analysis of selenate and selenite by single-column ion chromatography |
| |
Authors: | H. C. Mehra W. T. Frankenberger Jr. |
| |
Affiliation: | (1) Department of Soil & Environmental Sciences, University of California, 92521 Riverside, CA, USA;(2) Present address: University of Transkei, Transkei, Umtata, Southern Africa |
| |
Abstract: | ![]() Summary A rapid, selective and sensitive method has been developed for the simultaneous determination of selenate [Se(VI)] and selenite [Se(IV)] at trace levels by single-column ion chromatography. p-Hydroxybenzoic acid (4 mM) at pH 8.0 was used as the eluent. A low capacity resin-based anion-exchange column was employed for the separation with conductometric detection for quantification. The method requires minimal sample pretreatment. Resolution (Rs) between Se(VI)/Se(IV) was 6.76. Chloride, nitrate and sulfate did not interfere with selenium separation. The detection limits for Se(VI) and Se(IV) were 0.06 μg/ml and 0.110 μg/ml, respectively. The relative standard deviation using a 500-μl loop was 0.90 to 1.86% for Se(VI) and Se(IV), respectively. The method was applied to analyze seleniferous soil samples. |
| |
Keywords: | Ion chromatography Selenium speciation High speed ion analysis Conductometric detection |
本文献已被 SpringerLink 等数据库收录! |
|