首页 | 本学科首页   官方微博 | 高级检索  
     检索      


A specific sensitive HPLC method for determination of plasma dopamine
Authors:Parinam S Rao  Nisa Rujikarn  J M Luber Jr  D H Tyras
Institution:(1) Division of Cardiothoracic Surgery, Long Island Jewish Medical Center, 11042 New Hyde Park, NY, USA
Abstract:Summary We describe here a sensitive, selective and rapid method to quantitate plasma catecholamines, especially dopamine, using high-performance liquid chromatography with electrochemical detection. This method requires a 10-minute run time and has a threshold for detection of 2 picograms, (10pg/ml).A number of commonly employed mobile phases for catecholamine analysis have been tested and have failed to detect dopamine in biological samples. Neither acetonitrile (3–7%) or methanol, (5–8%) in the mobile phase has produced consistently interpretable data either due to inability to detect or interference from co-eluting substances. Optimal detection was achieved with a mobile phase containing sodium acetate (6.8g), citric acid (5.9g), EDTA (48mg), di-n-butylamine (270mgrl), Na-1-octane sulfate (850mg), methanol (100 ml) (amounts refer to 1 liter aqueous solution) (pH 4.3). The mobile phase was passed through a Waters 5mgr resolve C18 column using a Waters 590 pump and m460 electrochemical detector and 740 data module, Flow rate was 0.9ml/min. Using this method, normal values in human and swine left ventricular myocardium and human and swine plasma have been established for norepinephrine, epinephrine, and dopamine.
Keywords:Column liquid chromatography  Plasma Catecholamines  Dopamine  Electrochemical detection
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏