Analysis of long-term distributions of calibration parameters and calibration intervals for an atomic absorption spectrophotometer
Authors:I. Kuselman  Irena Goldfeld  Boris Anisimov
Affiliation:(1) The National Physical Laboratory of Israel (INPL), Givat Ram, Jerusalem 91904, Israel e-mail: kuselman@netvision.net.il Tel.: +972-2-653 6534 Fax: +972-2-652 797, IL
Abstract:
 A combination of "black box" and "calendar-time" methods for the determination of calibration intervals of an analytical measuring instrument is discussed. Since the methods require information on the distributions of the calibration parameters, such information is described for an atomic absorption spectrophotometer, as an example. The hypotheses on the normal distribution of the calibration parameters are tested using the ω2-criterion and accepted at 0.90–0.95 levels of confidence. Corresponding control charts are designed for indication of warning and action limits of the calibration parameters, and diagnoses of outliers in further calibrations. Control charts indicate also when the calibration should be done according to the full program of the equipment manufacturer. Received: 15 April 2000 / Accepted: 24 July 2000
Keywords:Calibration parameters  Measuring instrument  Spectrophotometry  Long-term distributions  Control charts  Calibration intervals