Automated determination of molybdenum(VI) in seawater by means of constant-current reduction of the adsorbed 8-quinolinol complex in a computerized flow potentiometric stripping analyzer |
| |
Authors: | Chi Hua Daniel Jagner Lars Renman |
| |
Affiliation: | Department of Technical Analytical Chemistry, Chemical Center, University of Lund, P.O. Box 124, S-221 00 Lund Sweden |
| |
Abstract: | Molybdenum(VI) in seawater is determined by means of potentiostatic adsorption of the 8-quinolinol complex onto a mercury film electrode at ?0.2 V vs. SCE and subsequent reduction of the complex by means of constant-current stripping in 5 M calcium chloride medium with a fully automated stripping analyzer. A single stripping peak at –0.42 V vs. SCE was obtained. The molybdenum(VI) concentration in reference seawater NASS-1, with a certified value of 11.5 ± 1.9 μg 1?1, was found to be 8.9 ± 1.3 μg 1?1 (n = 10). |
| |
Keywords: | |
本文献已被 ScienceDirect 等数据库收录! |