Ultraviolet absorbance detection of colchicine and related alkaloids on a capillary electrophoresis microchip |
| |
Authors: | Lu Qin Copper Christine L Collins Greg E |
| |
Affiliation: | a Naval Research Laboratory, 4555 Overlook Ave., S.W., Chemistry Division, Code 6112, Washington, DC 20375-5342, United States b United States Naval Academy, Chemistry Department, 572 Holloway Road, Annapolis, MD 21402, United States |
| |
Abstract: | The separation and UV absorbance detection of four toxic alkaloids, colchicine, thiocolchicine, colchicoside, and thiocolchicoside, on a microchip-based capillary electrophoresis device are reported. To increase the sensitivity of UV absorbance detection, optical cells with extended path lengths were integrated into the separation channel during the microfabrication process. The absorbance values realized on the microchip using these optical cells were proportional to the increase in average depths according to the Beer-Lambert Law, resulting in sensitivity enhancements by as much as five times. Linearity of response was observed from 5.0 to 500 mg L−1 of colchicine, with detection limits ranging from 2 to 6 mg L−1 depending upon the specific alkaloid and the dimension of the optical cell. The extraction of colchicine from spiked milk samples was performed and an average recovery rate of 83% with a relative standard deviation of 3.8% was determined using the optimized conditions on the microchip. |
| |
Keywords: | Colchicine Alkaloids Ultraviolet absorbance Microchip Capillary electrophoresis |
本文献已被 ScienceDirect PubMed 等数据库收录! |
|