首页 | 本学科首页   官方微博 | 高级检索  
     


k0 and Comparator NAA: Influences and Interactions
Authors:F. De Corte
Affiliation:(1) Laboratory of Analytical Chemistry, Institute for Nuclear Sciences, University of Gent, Proeftuinstraat 86, B-9000 Gent, Belgium
Abstract:
A discussion is held on mutual influences and interactions between k0- and (relative) comparator-type NAA. Examples are given concerning: (1) the application of comparator-type NAA in the quality control/quality assurance of the IRMM-530 Al-0.1%Au neutron flux monitor developed for use in k0-NAA, (2) the utilization of the k0-method of calibration as a tool for the quality assurance of comparator-type NAA, (3) the introduction of corrections for detection efficiency and true-coincidence (of primordial importance in k0-NAA) in comparator-type NAA, (4) the development of k0-type standardization for use in prompt gamma neutron activation analysis, and (5) the renewal of insights in the traceability of k0- and comparator-type NAA.
Keywords:
本文献已被 SpringerLink 等数据库收录!