首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Flow injection chemiluminescence determination of hydrazine by oxidation with chlorinated isocyanurates
Authors:Safavi Afsaneh  Karimi Mohammad Ali
Institution:Department of Chemistry, Faculty of Sciences, College of Sciences, Shiraz University, Shiraz 71454, Iran.
Abstract:A rapid and sensitive flow injection chemiluminescence (CL) method is described for the determination of hydrazine based on the CL generated during its reaction with either sodium dichloroisocyanurate (SDCC) or trichloroisocyanuric acid (TCCA) in alkaline medium. The emission intensity is greatly enhanced if dichlorofluorescein (DCF) as sensitizer is present in the reaction medium. The presence of citrate prevents the precipitation of some cations in the reaction medium and also causes an enhancement in emission intensity. The effect of analytical and flow injection variables on these CL systems and determination of hydrazine are discussed. The optimum parameters for the determination of hydrazine were studied and were found to be the following: SDCC and TCCA both 1x10(-3) M; NaOH, 2x10(-1) M; DCF, 5x10(-6) M; citrate, 1x10(-3) M and flow rate, 3.8 ml min(-1). The optimized method yielded 3sigma detection limits of 2x10(-7) and 3x10(-7) M for hydrazine with SDCC and TCCA oxidants, respectively. The method is simple, fast, sensitive, and precise and was applied to the determination of hydrazine in water samples.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏