首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Injection by Programmed Temperature Vaporization Injection (PTV) of Gaseous Samples for Gas Chromatography – Atomic Emission Spectrometry (GC-AED)
Authors:Suwannee Junyapoon  Andrew B Ross  Keith D Bartle  Bernard Frere  Alastair C Lewis  Michael Cooke
Abstract:Programmed temperature vaporization injection (PTV) coupled to gas chromatography and atomic emission detector (AED) has been studied for large volume injection of gaseous samples. As examples of the effectiveness of the technique, the results of the analysis of a series of headspace samples of foods such as garlic and onion, and of landfill gases are presented. The volumes of gaseous samples reconcentrated varied from a few milliliters up to liters depending on analyte dilution, through focusing onto a sorbent trap, then rapid liberation into the GC-AED system by programmed thermal desorption. Despite the high carrier gas flow rates associated with direct PTV-GC, AED performance and sensitivity were unaffected. The detailed elemental information obtained from the PTV-GC-AED analyses was confirmed using a PTV coupled to a gas chromatograph with ion trap detector mass spectrometer as detector (PTV-GC-ITD/MS).
Keywords:Programmed temperature vaporization (PTV)  GC-AED
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号