首页 | 本学科首页   官方微博 | 高级检索  
     


Research of reflectance photometer based on optical absorption
Authors:Lihua Huang  Youbao Zhang  Jianfeng Qu  Xiangzhao Wang
Affiliation:a Shanghai Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Shanghai 201800, China
b Graduate University of Chinese Academy of Sciences, Beijing 100049, China
Abstract:Gold-labeled lateral flow (LF) assay has been studied extensively due to its sensitiveness, simpleness, and rapidness. In this paper, scattering properties of nanogold particles and LF strip were analyzed. Based on the above analysis, a reflectance photometer has been developed for LF strip labeled by nanogold particles. The integration of reflection optical density was used to indicate the signal of T line and C line. A quantitative equation for cardiac troponin I (cTnI) was deduced to realize quantification of the real cTnI sample, whose correlation of coefficient (R2) was 0.989. The variation coefficient (CV) of repetitive experiments was less than 5%. The reflectance photometer was simple, sensitive, stable, and suitable for field detection.
Keywords:Nanogold particles   Lateral flow strip   Ideal diffuse reflector   Reflectance photometer
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号