首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of tyrosine traces by adsorption voltammetry of its copper (II) complex
Authors:Lizeng Wang   Chengsong Ma   Xiaoli Zhang   Yibing Ren  Yong Yu
Affiliation:(1) Department of Chemistry, Shandong University, 250100 Jinan, Shandong, People's Republic of China
Abstract:The electrochemical behavior of the copper-tyrosine complex has been studied by linear-sweep adsorption voltammetry. In 0.02 mol/L Na2HPO4 buffer solution (pH=9.6), the complex can be adsorped on a hanging mercury drop electrode and reduced at a peak potential of about –0.42 V (vs. SCE). The secondary derivative peak height is linear proportional to the concentration of tyrosine in the range 1.0×10–7–5.0×10–5 mol/L. The detection limit is 5×10–8 mol/L.Project supported by the Provincial Science Foundation of Shandong Province
Keywords:
本文献已被 SpringerLink 等数据库收录!