Dithizone–chloroform single drop microextraction system combined with electrothermal atomic absorption spectrometry using Ir as permanent modifier for the determination of Cd in water and biological samples |
| |
Affiliation: | 1. Grupo Química Analítica Aplicada (QANAP), University Institute of Research in Environmental Studies (IUMA), Centro de Investigaciones Científicas Avanzadas (CICA), University of A Coruña, A Coruña, Spain;2. University of Santiago de Compostela, Santiago de Compostela, Spain;1. School of Pharmacy, Guiyang Medical College, Guiyang 550004, China;2. MOE Key Laboratory of Aquatic Product Safety/KLGHEI of Environment and Energy Chemistry, School of Chemistry and Chemical Engineering, Sun Yat-sen University, Guangzhou 510275, China |
| |
Abstract: | A simple and sensitive method using dithizone–chloroform single drop microextraction has been developed for separation and preconcentration of trace Cd prior to its determination by electrothermal atomic absorption spectrometry with Ir as permanent modifier. Parameters, such as pyrolysis and atomization temperature, solvent type, pH, dithizone concentration, extraction time, organic drop volume, stirring rate and sample volume were investigated. Under the optimized conditions, a detection limit (3σ) of 0.7 ng/l and enrichment factor of 65 were achieved. The relative standard deviation was 7.4% (c = 0.2 μg/l, n = 5). The developed method has been applied to the determination of trace Cd in water samples and biological reference materials with satisfactory results. |
| |
Keywords: | |
本文献已被 ScienceDirect 等数据库收录! |
|