首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Efficiency of a miniaturized silica monolithic cartridge in reducing matrix ions as demonstrated in the simultaneous extraction of morphine and codeine from urine samples for quantification with liquid chromatography-tandem mass spectrometry (LC-MS/MS)
Authors:Nema T  Chan E C Y  Ho P C
Institution:Department of Pharmacy, National University of Singapore, 18 Science Drive 4, Singapore, 117543.
Abstract:Presence of matrix ions could negatively affect the sensitivity and selectivity of liquid chromatography‐tandem mass spectrometer (LC‐MS/MS). In this study, the efficiency of a miniaturized silica monolithic cartridge in reducing matrix ions was demonstrated in the simultaneous extraction of morphine and codeine from urine samples for quantification with LC‐MS. The miniaturized silica monolith with hydroxyl groups present on the largely exposed surface area function as a weak cation exchanger for solid phase extraction (SPE). The miniaturized silica cartridge in 1 cm diameter and 0.5 cm length was housed in a 2‐ml syringe fixed over a SPE vacuum manifold for extraction. The cleaning effectiveness of the cartridge was confirmed by osmometer, atomic absorption spectrometer, LC‐MS and GC‐TOFMS. The drugs were efficiently extracted from urine samples with recoveries ranging from 86% to 114%. The extracted analytes, after concentration and reconstitution, were quantified using LC‐MS/MS. The limits of detection for morphine and codeine were 2 ng/ml and 1 ng/mL, respectively. The relative standard deviations of measurements ranged from 3% to 12%. The monolithic sorbent offered good linearity with correlation coefficients > 0.99, over a concentration range of 50–500 ng/ml. The silica monolithic cartridge was found to be more robust than the particle‐based packed sorbent and also the commercial cartridge with regards to its recyclability and repeated usage with minimal loss in efficiency. Our study demonstrated the efficiency of the miniaturized silica monolith for removal of matrix ions and extraction of drugs of abuse in urinary screening. Copyright © 2011 John Wiley & Sons, Ltd.
Keywords:silica monolithic cartridge  solid phase extraction  morphine  codeine  urine  LC‐MS/MS
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号