首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Use of disposable open tubular ion exchange pre-columns for in-line clean-up of serum and plasma samples prior to capillary electrophoretic analysis of inorganic cations
Authors:Kiplagat Isaac K  Doan Thi Kieu Oanh  Kubáň Pavel  Kubáň Petr  Bo?ek Petr
Institution:Department of Chemistry and Biochemistry, Mendel University in Brno, Zemědělská 1, CZ-613 00 Brno, Czech Republic.
Abstract:A simple analytical system using disposable, open-tubular ion exchange clean-up precolumns coupled in-line to capillary electrophoresis for direct injection of biological samples is presented. The clean-up precolumns were prepared from fused silica capillaries by thermally initiated layer-by-layer polymerization of poly(butadiene-maleic acid) (PBMA) directly on the capillary wall. Typically, 6 cm long precolumns with 4-layers of PBMA were used for sample pretreatment. A robust and reproducible coupling between the precolumn (75 μm ID) and the analytical capillary (50 μm ID) was achieved using an inexpensive, commercially available low dead volume union. No extra dispersion of the analyte zones was observed. Proteins and other high molecular weight compounds from biological sample matrices were retained on the cation-exchanger sites of the precolumn, which eliminated their adsorption on analytical capillary walls and ensured stable electroosmotic flow and migration times of target analytes. Unretained small inorganic cations migrated freely into the analytical capillary for separation and detection. Applicability of the sample clean-up procedure was proved by determination of major inorganic cations in blood serum and plasma samples using capillary electrophoresis with contactless conductivity detection. Separations were performed in background electrolyte solution consisting of 15 mM L-arginine, 12.5 mM maleic acid, 3 mM 18-crown-6 at pH 5.5 and repeatabilities of migration times and peak areas were below 1.5% and 7.3%, respectively. Less than 1 μL of biological sample was required for injection.
Keywords:Biological samples  Capillary electrophoresis  Capacitively coupled contactless conductivity detection  On-line sample clean-up  Open-tubular ion exchange columns
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号