首页 | 本学科首页   官方微博 | 高级检索  
     


On-line determination of mercury(II) by membrane separation flow injection analysis
Authors:Amini Nazanin  Cardwell Terence J  Cattrall Robert W  Morrison Richard J S  Kolev Spas D
Affiliation:a School of Chemistry, The University of Melbourne, Parkville, Vic. 3010, Australia
b Department of Chemistry, Analytical Chemistry Laboratories, La Trobe University, Melbourne, Vic. 3086, Australia
c School of Chemistry, Monash University, Melbourne, Vic. 3800, Australia
Abstract:A rapid and inexpensive gas-diffusion (GD) flow injection method for the on-line determination of Hg(II) in aqueous samples is described. The analytical procedure involves the injection of a Hg(II) sample into a 1.5 M H2SO4 carrier stream which is merged with a reagent stream containing 0.6% SnCl2 and 1.5 M H2SO4. Under these conditions Hg(II) is reduced to metallic mercury which partially evaporates through a Teflon membrane into an acceptor stream containing 1.75×10−4 M KMnO4 in 0.3 M H2SO4. The decrease in the absorbance of the acceptor stream at 528 nm corresponding to the absorption maximum of the permanganate anion can be related to the original concentration of Hg(II) in the sample. The method is characterized by a detection limit of 4 μg l−1 and a sampling frequency of 8 h−1. The flow system was successfully applied to the analysis of river samples spiked with Hg(II).
Keywords:Gas-diffusion   Flow injection   Mercury determination   Cold vapor
本文献已被 ScienceDirect PubMed 等数据库收录!