Trace enrichment and determination of silver by immobilized DDTC microcolumn and flow injection atomic absorption spectrometry |
| |
Authors: | Dadfarnia S Haji Shabani A M Gohari M |
| |
Affiliation: | Department of Chemistry, Yazd University, Yazd 89195-741, Iran |
| |
Abstract: | ![]() Flow injection (FI) system incorporating a microcolumn of immobilized diethyldithiocarbamate (DDTC) on surfactant-coated alumina was combined with atomic absorption spectrometry for on-line trace enrichment and determination of silver in different matrices. Silver was deposited on the microcolumn by processing a standard or solution of analyte at pH 3-4 on the column. Injection of 250 μl of ethanol then served to elute the retained species to atomic absorption spectrometry (AAS). A sample volume of 20 ml resulted in a pre-concentration factor of 125, and precision at the 20 μg l−1 was 4% (R.S.D.). The procedure was applied to tap water, well water, rain water, sea water, radiology film, and lead concentrate samples. The accuracy was assessed through recovery experiments, independent analysis by furnace-AAS, and analysis of certified reference material. |
| |
Keywords: | Silver determination Immobilized diethyldithiocarbamate On-line pre-concentration Flow injection |
本文献已被 ScienceDirect PubMed 等数据库收录! |
|