Validation of bismuth film electrode for determination of cobalt and cadmium in soil extracts using ICP-MS |
| |
Authors: | Hutton Emily A van Elteren Johannes T Ogorevc Bozidar Smyth Malcolm R |
| |
Institution: | a Analytical Chemistry Laboratory, National Institute of Chemistry, Hajdrihova 19, SI-1000 Ljubljana, Slovenia b National Centre for Sensor Research, Dublin City University, Glasnevin, Dublin 9, Ireland |
| |
Abstract: | A study is presented on the use of the bismuth film electrode (BiFE) operated in the anodic stripping and the cathodic adsorptive stripping voltammetry (ASV, CAdSV) modes, for the determination of two trace heavy metals (Cd and Co, respectively), in soil extract samples. Two types of BiFE were examined in this study: the in situ prepared BiFE, which was employed in ASV determination of Cd, and the ex situ prepared BiFE, which was used in CAdSV of Co with dimethylglyoxime (DMG) as complexing agent. A series of unpretreated soil extracts with varying Cd and Co concentrations were analyzed, and the results obtained compared to those determined using inductively coupled plasma-mass spectrometry (ICP-MS). The results revealed the suitability of stripping analysis at the BiFE for determination of μg l−1 levels of heavy metals in soil extracts. The promising results obtained here, coupled with the non-toxic nature of bismuth (in comparison to commonly used mercury electrodes employed in stripping analysis), offer great promise in centralized and decentralized analysis of trace heavy metals in complex environmental matrices. |
| |
Keywords: | Bismuth film electrode Stripping analysis Cadmium Cobalt Soil extract ICP-MS Heavy metal |
本文献已被 ScienceDirect PubMed 等数据库收录! |
|