首页 | 本学科首页   官方微博 | 高级检索  
     


A phase-sensitive technique to measure small changes in laser frequency: application to measure the shift and broadening of a saturated absorption line of OsO4
Authors:K. J. Siemsen  J. E. Bernard
Affiliation:(1) Institute for National Measurement Standards, National Research Council, Ottawa, K1A 0R6, Canada
Abstract:
A phase-sensitive technique was used to measure the sensitivity to operating conditions of the output frequency of a highly stable carbon dioxide laser, which operated on the 10-μmm R(0) transition (28.8 THz) and was locked to the saturated absorption of the Q(15) transition of OsO4. Periodic changes in the operating conditions were imposed on the laser and the resulting changes in the output frequency were measured through phase-sensitive demodulation of the heterodyne beat with a second CO2 laser of much lower stability. Shifts as small as 50Hz could be detected by this technique. The results are in good agreement with previous measurements which employed an infrared frequency chain.
Keywords:42.62.Eh  42.62.Fi  33.20.Ea
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏