Mass selective separation applied to radioisotopes of cesium
Authors:Michael Dion  Greg Eiden  Orville Farmer  Zach Finch  Martin Liezers
Institution:Pacific Northwest National Laboratory, Richland, WA, USA
Abstract:A technique that uses the intrinsic mass‐based separation capability of a quadrupole mass spectrometer has been used to resolve spectral radiometric interference of two isotopes of the same element. In this work the starting sample was a mixture of 137Cs and 134Cs and was (activity) dominated by 137Cs. This methodology separated and ‘implanted’ 134Cs that was later quantified for spectral features and activity with traditional radiometric techniques. This work demonstrated a 134Cs/137Cs activity ratio enhancement of >4 orders of magnitude and complete removal of 137Cs spectral features from the implanted target mass (i.e. 134). Copyright © 2016 John Wiley & Sons, Ltd.
Keywords:icp‐ms  ion implantation  radiometric quantification  radioisotopes  isotope separations