首页 | 本学科首页   官方微博 | 高级检索  
     


Evaluation of packed capillary liquid chromatography columns and comparison with conventional-size columns
Authors:Prüss Anja  Kempter Christine  Gysler Jens  Jira Thomas
Affiliation:Schering AG, Analytical Development 3, Müllerstr 178, D-13342 Berlin, Germany.
Abstract:
Apart from extracolumn effects peak dispersion in liquid chromatographic columns is caused by the column inlet, the packed bed, and the column outlet. A strategy applicable for independent evaluation of the individual sources of column band broadening was developed on the basis of the linear extrapolation method (LEM). This method was applied to compare the performance of packed capillary LC columns from various commercial suppliers with conventional-size columns. The columns differed widely in their performance with respect to peak shapes and widths for standard substances. The capillary columns were found well packed, but in some cases overall performance would benefit from improving the design of the area between the packed bed and the connecting capillaries, containing frits as well as dead volumes.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏