Comparison of the column performance of narrow-bore and standard-bore columns for the chromatographic determination of alpha-, beta-, gamma-, and delta-tocopherol |
| |
Authors: | Ye L Landen W O Eitenmiller R R |
| |
Affiliation: | Department of Food Science and Technology, University of Georgia, Athens 30602, USA. |
| |
Abstract: | A comparison of the performance of narrow-bore (2.1-mm i.d.) and standard-bore (4.6-mm i.d.) analytical silica columns having the same length is completed for the resolution of alpha-, beta-, gamma-, and delta-tocopherol. The studies are performed on high-performance liquid chromatographic equipment with minimum extracolumn contribution. Column permeabilities are 1.16 x 10(-9) and 2.48 x 10(-9) cm2 for narrow and standard bore, respectively. The narrow-bore column gives up to a 7 times increase in sensitivity compared with a standard-bore column at equivalent running times for the analytes. Approximately one-third solvent savings can be achieved with the narrow-bore column. Theoretical plates of the standard-bore column are higher than that of the narrow-bore column. |
| |