Utilization of a Novel Immunofluorescence Instrument Prototype for the Determination of the Herbicide Glyphosate |
| |
Authors: | Eszter Taká cs,Borbá la Gé mes,Fanni Szendrei,Csaba Keszei,Attila Baró csi,Sá ndor Lenk,Lá szló Domjá n,Má ria Mö rtl,Andrá s Szé ká cs |
| |
Abstract: | An enzyme-linked fluorescent immunoassay (ELFIA) method has been developed for the quantitative analytical determination of the herbicide active ingredient glyphosate in environmental matrices (surface water, soil, and plant tissues). Glyphosate, as a ubiquitous agricultural pollutant, is a xenobiotic substance with exposure in aquatic and terrestrial ecosystems due its extremely high worldwide application rate. The immunoassay developed in Project Aquafluosense is part of a fluorescence-based instrumentation setup for the in situ determination of several characteristic water quality parameters. The 96-well microplate-based competitive immunoassay method applies fluorescence signal detection in the concentration range of 0–100 ng/mL glyphosate. Application of the fluorescent signal provides a limit of detection of 0.09 ng/mL, which is 2.5-fold lower than that obtained with a visual absorbance signal. Beside the improved limit of detection, determination by fluorescence provided a wider and steeper dynamic range for glyphosate detection. No matrix effect appeared for the undiluted surface water samples, while plant tissues and soil samples required dilution rates of 1:10 and 1:100, respectively. No cross-reaction was determined with the main metabolite of glyphosate, N-aminomethylphosphonic acid, and related compounds. |
| |
Keywords: | glyphosate competitive immunoassay ELFIA fluorescence detection Project Aquafluosense |
|
|