Abstract: | ![]() A simple micellar electrokinetic chromatographic method is described for simultaneous determination of digoxin and digitoxin. The simultaneous analysis of digoxin and digitoxin was performed in Tris buffer (10 mM; pH 9) with 90 mM sodium dodecyl sulfate and 10% isopropyl alcohol as an anionic surfactant and organic modifier. Under these conditions, good separation with high efficiency is achieved in short analysis times. Several parameters affecting the separation of the drugs were studied, including the pH and concentrations of the Tris buffer and sodium dodecyl sulfate. The linear range of the method for the determination of digoxin and digitoxin was over 0.01–0.3 mg/mL; the detection limit (signal to noise ratio = 3; injection 3.5 kPa 3 s) was 4 and 6 μg/mL, respectively. Application of the proposed method to the determination of digoxin in commercial tablets and in injections proved to be feasible. |