Double-ion imprinted polymer @magnetic nanoparticles modified screen printed carbon electrode for simultaneous analysis of cerium and gadolinium ions |
| |
Authors: | Bhim Bali Prasad Darshika Jauhari |
| |
Institution: | Analytical Division, Department of Chemistry, Faculty of Science, Banaras Hindu University, Varanasi 221005, India |
| |
Abstract: | A typical, reproducible, and rugged screen printed carbon electrode, modified with dual-ion imprinted beads, was fabricated employing the “surface grafting from” approach. For this, the acyl chloride functionalized magnetic nanoparticles were first immobilized and chemically attached with a typical functional monomer (but-2-enedioic acid bis-(2-amino-ethyl)-amide]) on the electrode surface. This was subsequently subjected to the thermal polymerization in the presence of template ions (Ce(IV) and Gd(III)), cross-linker (ethylene glycol dimethacrylate), initiator (AIBN), and multiwalled carbon nanotubes. The modified sensor was used for the simultaneous analysis of both template ions in aqueous, blood serum, and waste-water samples, using differential pulse anodic stripping voltammetry which revealed two oxidation peaks for respective templates with resolution as much as 950 mV, without any cross reactivity, interferences and false-positives. The detection limits realized by the proposed sensor, under optimized conditions, were found to be as low as 0.07 ng mL−1 for Ce(IV) and 0.19 ng mL−1 for Gd(III) (S/N = 3) that could eventually be helpful for lanthanide estimation at stringent levels. |
| |
Keywords: | Dual ion imprinted polymer Screen printed carbon electrode Lanthanide ions Differential pulse anodic stripping voltammetry Simultaneous ultra trace analysis |
本文献已被 ScienceDirect 等数据库收录! |
|