Model for the analytical signal in atomic absorption spectrometry with electrothermal atomization |
| |
Authors: | Yu. V. Rogulsky A. N. Buhay A. A. Gudakova A. N. Kulik |
| |
Affiliation: | (1) Institute of Applied Physics, National Academy of Sciences of Ukraine, 58 Petropavlovskaya St., Sumy, 40030, Ukraine |
| |
Abstract: | A method for verification of analytical measurements in electrothermal atomic absorption spectrometry using a phenomenological model for physicochemical atomization processes is discussed. The influence of the sample matrix on basic kinetic parameters of atomization such as the analyte free-atom formation energy and the atomization rate constant has been shown. The similarity of atomization kinetic parameters obtained by calibrating the spectrometer to those for samples under study can be used as a verification criterion for analysis. __________ Translated from Zhurnal Prikladnoi Spektroskopii, Vol. 75, No. 3, pp. 295–298, May–June, 2008. |
| |
Keywords: | atomic absorption spectrometry simulation electrothermal atomization graphite furnace formation energy of free atoms kinetic parameters |
本文献已被 SpringerLink 等数据库收录! |
|