首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Modification of a packed column gas chromatograph/mass spectrometer
Authors:Tan Y L  Quanci J F
Institution:Environmental Measurements Laboratory, U.S. Department of Energy, New York, New York 10014, U.S.A.
Abstract:A packed-column gas-chromatograph/mass-spectrometer (GC/MS), Hewlett-Packard 5982, was modified to accommodate fused silica capillary columns. The original GC/MS interface and chemical-ionization sample-line in the ion-source were changed to allow the end of a fused silica capillary column to enter the ion-chamber directly. For chemical-ionization operation, the reagent gas was brought into the MS through the direct-insertion probe port. The calibration compound was introduced through the electron-impact sample-inlet, which simplified the operation. The modified system yields higher sensitivity and more efficient separation, as well as simpler operation, without sacrificing any original instrument functions.
Keywords:Author for correspondence  
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号