首页 | 本学科首页   官方微博 | 高级检索  
     


The use of a micro-bubble agent to enhance rabbit liver destruction using high intensity focused ultrasound
Authors:Yu Tinghe  Xiong Shuhua  Mason Timothy J  Wang Zhibiao
Affiliation:Institute of Ultrasound Engineering in Medicine, Chongqing Medical University, Chongqing 400016, PR China. yutinghe@hotmail.com
Abstract:
Liver tissues in New Zealand rabbits were ablated using high intensity focused ultrasound (HIFU, 14300 W/cm(2), 1.0 MHz). The animals were intravenously administered 0.2 ml of micro-bubble agent in the experimental (n=20) group and an isovolumetric normal saline solution in the control (n=27) group before HIFU treatment which was performed as a linear scan. In both groups, the preselected tissue volumes were destroyed without harming the overlying tissues. Necrosis rate (NR, cm(3)/s) was used to reflect the ablation efficiency, which was the tissue volume of occurring coagulative necrosis per 1s HIFU exposure. NR was improved in the experimental group (0.0570+/-0.0433 vs 0.0120+/-0.0122, P=0.0002). Pathological studies confirmed that there were no residual intact targets within the exposed volume. These findings suggested that the introduction of the micro-bubble agent enhanced HIFU liver destruction.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号