首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Determination of As(III) and total as in water by graphite furnace atomic absorption spectrometry after electrochemical preconcentration on a gold-plated porous glassy carbon electrode
Authors:M St?elec  F ?acho  A Manová  E Beinrohr
Institution:(1) Faculty of Chemical and Food Technology, Institute of Analytical Chemistry, Slovak Technical University, Radlinského 9, 812 37 Bratislava, Slovakia
Abstract:Arsenic(III) was preconcentrated in a flow-through electrochemical cell on a gold coated porous carbon electrode. On stripping, arsenic was eluted with diluted nitric acid and determined off-line by GF AAS. The deposition and stripping steps were optimized. The limit of detection and limit of quantification were found to be 1.9 μg L1 and 6.4 μg L−1, respectively. The repeatability and reproducibility were found to be 5.3 % and 9.3 %, respectively. Total arsenic was determined after a microwave assisted chemical reduction of As(V) to As(III) making the procedure suitable for speciation analysis. The method was applied in analysis of water samples.
Keywords:electrochemical preconcentration  speciation  arsenic  flow-through cell  GF AAS
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号