Tamoxifen monitoring studies in breast cancer patients by micellar liquid chromatography
Authors:Josep Esteve-Romero  Enrique Ochoa-Aranda  Devasish Bose  Maria Rambla-Alegre  Juan Peris-Vicente  Adrià Martinavarro-Domínguez
Affiliation:1. Química Bioanalítica, QFA, ESTCE, Campus Riu Sec, Universitat Jaume I, 12071, Castellón de la Plana, Spain
2. Hospital Provincial, Dr. Clara, 12003, Castellón de la Plana, Spain
3. Criminology and Forensic Science, Dr. H.S. Gour University, Sagar, 470 003, MP, India
Abstract:
A simple micellar liquid chromatographic procedure is described to determine tamoxifen in plasma. To perform the analysis, tamoxifen solutions were diluted in water and UV-irradiated for 20 min to form the photocycled derivative with a phenanthrene core which shows intense fluorescence. Samples were then directly injected, thus avoiding long extraction and experimental procedures. The resolution from the matrix was performed with a mobile phase containing 0.15 M SDS–7% n-butanol at pH 3 running at 1.5 mL/min through a C18 column at 40 °C. Detection was carried out by fluorescence, and the excitation and emission wavelengths were 260 and 380 nm, respectively. The chromatographic analysis time was less than 15 min. The analytical methodology was validated following the International Conference on Harmonisation of Technical Requirements for Registration of Pharmaceuticals for Human Use (ICH) guidelines. The response of the drug in plasma was linear and in the 0.5–15 μg/mL range, with r 2 > 0.999. Accuracy and precision were <9% in both cases. The limits of detection and quantification (in nanograms per millilitre) were 50 and 150 in plasma, respectively. The method developed herein shows no interferences by endogenous compounds. Finally, the analytical method was used to determine the amount of tamoxifen in the plasma of several breast cancer patients from a local hospital.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号