首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Separation and detection of amino acid metabolites of Escherichia coli in microbial fuel cell with CE
Authors:Wei Wang  Lihong Ma  Ping Lin  Kaixuan Xu
Institution:Ministry of Education Key Lab of Analysis and Detection for Food Safety, Fujian Provincial Key Lab of Analysis and Detection for Food Safety, School of Chemistry, Fuzhou University, Fuzhou, Fujian, P. R. China
Abstract:In this work, CE‐LIF was employed to investigate the amino acid metabolites produced by Escherichia coli (E. coli) in microbial fuel cell (MFC). Two peptides, l ‐carnosine and l ‐alanyl‐glycine, together with six amino acids, cystine, alanine, lysine, methionine, tyrosine, arginine were separated and detected in advance by a CE‐LIF system coupled with a homemade spontaneous injection device. The injection device was devised to alleviate the effect of electrical discrimination for analytes during sample injection. All analytes could be completely separated within 8 min with detection limits of 20–300 nmol/L. Then this method was applied to analyze the substrate solution containing amino acid metabolites produced by E. coli. l ‐carnosine, l ‐alanyl‐glycine, and cystine were used as the carbon, nitrogen, and sulfur source for the E. coli culture in the MFC to investigate the amino acid metabolites during metabolism. Two MFCs were used to compare the activity of metabolism of the bacteria. In the sample collected at the running time 200 h of MFC, the amino acid methionine was discovered as the metabolite with the concentrations 23.3 μg/L.
Keywords:Capillary electrophoresis  Escherichia coli  Laser‐induced fluorescence  Metabolites
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号