Ionic liquid-based single drop microextraction combined with electrothermal atomic absorption spectrometry for the determination of manganese in water samples
Authors:Jamshid L Manzoori  Mohammad Amjadi  Jafar Abulhassani
Institution:Department of Analytical Chemistry, Faculty of Chemistry, University of Tabriz, Tabriz, Iran
Abstract:Room temperature ionic liquid, 1-butyl-3-methylimidazolium hexafluorophosphate C4MIM]PF6], was used as an alternative to volatile organic solvents in single drop microextraction-electrothermal atomic absorption spectrometry (SDME-ETAAS). Manganese was extracted from aqueous solution into a drop of the ionic liquid after complextaion with 1-(2-thiazolylazo)-2-naphthol (TAN) and the drop was directly injected into the graphite furnace. Several variables affecting microextraction efficiency and ETAAS signal, such as pyrolysis and atomization temperature, pH, TAN concentration, extraction time, drop volume and stirring rate were investigated and optimized. In the optimum experimental conditions, the limit of detection (3 s) and the enhancement factor were 0.024 μg L−1 and 30.3, respectively. The relative standard deviation (RSD) for five replicate determinations of 0.5 μg L−1 Mn(II) was 5.5%. The developed method was validated by the analysis of a certified reference material (NIST SRM 1643e) and applied successfully to the determination of manganese in several natural water samples.
Keywords:Manganese  Ionic liquid  Single drop microextraction  Electrothermal atomic absorption spectrometry  1-(2-thiazolylazo)-2-naphthol