A high‐performance liquid chromatography–tandem mass spectrometry method coupled with protein precipitation for determination of granisetron in human plasma and its application to a comparative pharmacokinetic study |
| |
Authors: | Ying Zhou Ji Jiang Pei Hu Hongyun Wang |
| |
Affiliation: | Clinical Pharmacology Research Center, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China |
| |
Abstract: | A rapid, simple and validated method based on liquid chromatography coupled with tandem mass spectrometry (LC‐MS/MS) has been developed for the determination of granisetron in human plasma. Plasma samples were pre‐purified by protein precipitation procedure. The chromatographic separation was achieved with Synergi Polar‐RP (75 × 2 mm, 4 µm) column using a mixture of 5 mm pH4.0 ammonium formate and methanol (300:316, v/v) under isocratic conditions at a flow rate of 0.3 mL/min. The detection was performed on a triple quadrupole tandem mass spectrometer in multiple reaction monitoring mode using positive electrospray ionization. The analysis time was about 2.5 min. The method was fully validated over the concentration range 0.1–10 ng/mL. The lower limit of quantification was 0.1 ng/mL. Inter‐ and intra‐batch precision was <6.1% and the accuracy was within 95.6–100.0%. The mean extraction recovery was 96.3%. Selectivity, matrix effect and stability were also validated. The method was applied to the comparative pharmacokinetic study of granisetron in Chinese healthy subjects. Copyright © 2014 John Wiley & Sons, Ltd. |
| |
Keywords: | determination granisetron LC‐MS/MS human plasma pharmacokinetic |
|
|