首页 | 本学科首页   官方微博 | 高级检索  
     


Anodic stripping voltammetric determination of trace lead with a nafion/crown-ether film electrode
Authors:Dong S  Wang Y
Affiliation:Changchun Institute of Applied Chemistry, Academia Sinica, Changchun, Jilin 130021, People's Republic of China.
Abstract:An electrode for the anodic stripping voltammetric determination of trace lead has been made by coating glassy carbon with a film of Naflon in which a crown ether is incorporated. The sensitivity is increased because of continuous transfer of lead from solution to the electrode surface by complexation with the crown ether and reduction during the deposition period. A detection limit of 5 x 10(-10)M has been obtained by electrodeposition for 3 min. The method is sensitive, simple and relatively rapid, with a relative standard deviation of 8% at the 2 x 10(-9)M level.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号