首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Determination of fluorine and chlorine in geological materials by induction furnace pyrohydrolysis and standard-addition ion-selective electrode measurement
Authors:Rice T D
Institution:Mineral Resources Development Laboratory, New South Wales Department of Mineral Resources, Post Office Box 76, Lidcombe, 2141, Australia.
Abstract:Fluorine and chlorine in geological materials are volatilized by pyrohydrolysis at about 1150 degrees in a stream of oxygen (1000 ml/min) plus steam in an induction furnace. The catalyst is a 7:2:1 mixture of silica gel, tungstic oxide and potassium dihydrogen phosphate. The sample/catalyst mixture is pyrohydrolysed in a re-usable alumina crucible (already containing four drops of 1 + 3 phosphoric acid) inserted in a silica-enclosed graphite crucible. The absorption solution is buffered at pH 6.5 and spiked with 1.6 mug of fluoride and 16 mug of chloride per g of solution, to ensure rapid and linear electrode response during subsequent standard-addition measurement. The simple plastic absorption vessel has 99.5% efficiency. The 3s limits of detection are 5-10 mug/g and 40-100 mug/g for fluorine and chlorine respectively. The procedure is unsuitable for determining chlorine in coal.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏