Using calibration approaches to compensate for remaining matrix effects in quantitative liquid chromatography/electrospray ionization multistage mass spectrometric analysis of phytoestrogens in aqueous environmental samples
Authors:Kang Jinguo  Hick Larry A  Price William E
Institution:School of Chemistry, University of Wollongong, Wollongong, NSW 2522, Australia.
Abstract:Signal suppression is a common problem in quantitative liquid chromatography/electrospray ionization multistage mass spectrometric (LC/ESI-MS(n)) analysis in environment samples, especially in highly loaded wastewater samples with highly complex matrix. Optimization of sample preparation and improvement of chromatographic separation are prerequisite to improve reproducibility and selectivity. Matrix components are reduced if not eliminated by optimization of sample preparation steps. However, extensive sample preparation may be time-consuming and risk the significant loss of some trace analytes. The best way to further compensate matrix effects is the use of an internal standard for each analyte. However, in a multi-component analysis, finding appropriate internal standards for every analyte is often difficult. In this present study, a more practical alternative option was sought. Matrix effects were assessed using the post-extraction addition method. By comparison of three different calibration approaches, it was found that matrix-matched calibration combined with one internal standard provides a satisfactory method for compensating for any residual matrix effects on all the analytes. Validating experiments on different sewage treatment plant (STP) influent samples analyzing for a range of phytoestrogens showed that this calibration method provided satisfactory results with concentration ratio 96.1-105.7% compared to those by standard addition.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号