A procedure is described in which a wide bore glass capillary column is used as an alternative to the more traditional packed column in the analysis of amino acid levels in plasma. The coefficients of variation for all amino acids (with the exception of aspartic acid) were better than 11% with recoveries ranging from 81% to 122%. The data are compared with the corresponding results obtained using a packed column and show significant differences (p < 0.005) between values for glycine, serine, isoleucine, proline, methionine, aspartic acid, phenylalanine, and lysine. A similar comparison between results from the wide bore and the fused-silica open tubular (FSOT) column shows better agreement. Adjustment of chromatographic conditions for the wide bore analysis yields results in good agreement with those from FSOT analysis but which still differe significantly from the backed column data.
Keywords:Gas chromatography  Fused-silica capillary  Wide bore glass capillary  Packed column  Plasma amino acids