首页 | 本学科首页   官方微博 | 高级检索  
     检索      


DSC Calibration During Cooling
Authors:G Hakvoort  C M Hol
Institution:(1) Chemistry Department, Delft University of Technology, Julianalaan 136, 2628 BL Delft, The Netherlands
Abstract:A number of compounds are investigated for DSC calibration during cooling. Adamantane and Zn show fast reversible transitions and can be applied both for temperature and for heat calibration. A third compound, namely 4,4′-azoxyanisole, has a liquid crystal to isotropic liquid transition at 409 K. This compound can be used for temperature calibration. Heat calibration with this compound is more problematic because of the small heat effect and the construction of the baseline. Two other compounds, namely Hg and Pb, show a slight undercooling. Nevertheless they can be used for heat calibration, and possibly also for temperature calibration during cooling. This revised version was published online in August 2006 with corrections to the Cover Date.
Keywords:adamantane (C10H16)  avoidance of sample/pan reactions  4  4′  -azoxyanisole (C14H14N2O3)  DSC calibration during cooling  Ga  Pb  Hg  reversible transitions  undercooling  Zn
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏