首页 | 本学科首页   官方微博 | 高级检索  
     检索      


A selective modified carbon paste electrode for determination of cyanide using tetra-3,4-pyridinoporphyrazinatocobalt(II)
Authors:Abbaspour A  Asadi M  Ghaffarinejad A  Safaei E
Institution:Department of Chemistry, Shiraz University, Shiraz 71454, Iran
Abstract:A chemically modified carbon paste electrode with 3,4-tetra pyridinoporphirazinatocobalt(II) (Co(3,4 tppa) was applied to the determination of free cyanide ion. The electrode has a linear range between 1.5 × 10−5 M and 1.0 × 10−2 M with a Nernstian slope of 60 ± 1.5 mV/decade and its detection limit is 9 × 10−6 M. The response time of electrode is 5 min. The proposed electrode was applied successfully for the determination of cyanide in commercially available spring water. Some anions, such as SCN, I, Cl, Br and oxalate that are usually serious interfering species for most of cyanide selective electrodes, did not have any interfering effect for this proposed electrode.
Keywords:Carbon paste electrode  Cyanide  Potentiometry  Co(3  4 tppa)
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏