首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Determination of selected neurotransmitter metabolites using monolithic column chromatography coupled with chemiluminescence detection
Authors:Adcock Jacqui L  Barnett Neil W  Costin Jason W  Francis Paul S  Lewis Simon W
Institution:School of Biological and Chemical Sciences, Deakin University, Geelong, Vic. 3217, Australia
Abstract:The oxidation of selected clinically important neurotransmitter metabolites with acidic potassium permanganate in the presence of polyphosphates evokes chemiluminescence of sufficient intensity to enable the sensitive determination of these species. Limits of detection for 5-hydroxyindole-3-acetic acid (5-HIAA), vanilmandelic acid (VMA; α,4-dihydroxy-3-methoxybenzeneacetic acid), 4-hydroxy-3-methoxyphenylglycol (MHPG), homovanillic acid (HVA, 4-hydroxy-3-methoxyphenylacetic acid) and 3,4-dihydroxyphenylacetic acid (DOPAC) were between 5 × 10−9 and 4 × 10−8 M, using flow-injection analysis methodology. In addition, we demonstrate the rapid determination of homovanillic acid and 5-hydroxyindole-3-acetic acid in human urine - without the need for extraction procedures - using monolithic column chromatography with chemiluminescence detection.
Keywords:Neurotransmitter metabolites  Monolithic column chromatography  Chemiluminescence detection
本文献已被 ScienceDirect PubMed 等数据库收录!