首页 | 本学科首页   官方微博 | 高级检索  
     


MCM-41 mesoporous material modified carbon paste electrode for the determination of cardiac troponin I by anodic stripping voltammetry
Authors:Guo Huishi  He Nongyue  Ge Shuxun  Yang Di  Zhang Jinan
Affiliation:a State Key Laboratory of Bioelectronics (Chien-Shiung Wu Laboratory), Southeast University, Nanjing 210096, China
b Institute of Cardiovascular Disease, Nanjing Medical University, Nanjing 210029, China
c Laboratory of NanoBio Medicines and Devices, Zhuzhou Institute of Technology, Zhuzhou 412008, China
Abstract:
An anodic stripping voltammetric method for the determination of cardiac troponin I (cTnI) at a MCM-41 mesoporous material modified carbon paste electrode (MCM-MCPE) was investigated. The test was based on the dual monoclonal antibody “sandwich” principle using colloidal gold as a labeled substrate. Four main steps were carried out to obtain the analytical signal, i.e. electrode preparation, immunoreaction, silver enhancement, and anodic stripping voltammetric detection. The anodic stripping peak current increased linearly with the concentration of cTnI over the range of 0.8-5.0 ng/ml. A detection limit of 0.5 ng/ml was obtained. The established method was applied to detect cTnI in acute myocardial infarction (AMI) samples using routine enzyme-linked immunoadsorbent assay (ELISA) for comparison analysis, and good results were obtained.
Keywords:Cardiac troponin I   Mesoporous materials   MCM-41   Carbon paste electrode   Anodic stripping voltammetry
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏