首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of total antimony(III,V) by square-wave anodic stripping voltammetry with in situ plated bismuth-film electrode
Authors:Ping Zong  Jiajie Long
Affiliation:Department of Applied Chemistry and Biotechnology , Graduate School of Engineering, University of Fukui , Bunkyo, Fukui, 910-8507, Japan
Abstract:A sensitive and reliable method is described for the determination of total Sb(III,?V) at traces levels by Osteryoung square-wave anodic stripping voltammery (OSWASV). This method is based on the co-deposition of Sb(III,?V) with Bi(III) onto an edge-plane pyrolytic graphite substrate at an accumulation step. OSWASV studies indicated that the co-deposited antimony was oxidised with anodic scans to give an enhanced anodic peak at about 450?mV vs. Ag/AgCl (sat. KCl). The anodic stripping peak current was directly proportional to the total concentration of antimony in the ranges of 0.01–0.10?µg?L?1, 0.10–1.0?µg?L?1 and 1.0–18.0?µg?L?1 with correlation coefficient higher than 0.995 when 2.0?mol?L?1 hydrochloric acid was used. The detection limits calculated as S/N?=?3 was 5.0?ng?L?1 in 2.0?mol?L?1 hydrochloric acid at 180?s deposition time. The relative standard deviation was 5% (n?=?6) at 0.10?µg?L?1 level of antimony. The analytical results demonstrate that the proposed method is applicable to analyses of real water samples.
Keywords:antimony determination  in situ plating  bismuth film electrode  edge-plane pyrolytic graphite  OSWASV