Radio frequency field intensity mapping using a composite spin-echo sequence |
| |
Authors: | C H Oh S K Hilal Z H Cho I K Mun |
| |
Affiliation: | Department of Radiology, Columbia University, New York, NY 10032. |
| |
Abstract: | A novel radio frequency (RF) field intensity mapping or imaging method using a composite NMR spin-echo sequence is proposed. A composite spin-echo RF pulse with 90 degrees y-180 degrees x-90 degrees y sequence makes phase change in the final image depending on the RF field intensity on the object. The resultant phase change or phase map can be used to obtain the actual RF flip-angle map for a given condition which includes the status of tuning and RF inhomogeneity, etc. Bloch equation has been solved numerically to obtain the effects of the RF field intensity as well as the main magnetic field inhomogeneity and the results are used for the mapping (imaging) of the RF field intensity. Phantom studies have been performed using a 1.5 Tesla whole body MRI system and the results are presented. |
| |
Keywords: | |
|
|