Quantification of grafted poly(ethylene glycol)-silanes on silicon by time-of-flight secondary ion mass spectrometry |
| |
Authors: | Norrman K Papra A Kamounah F S Gadegaard N Larsen N B |
| |
Affiliation: | Danish Polymer Centre, Ris? National Laboratory, DK-4000 Roskilde, Denmark. kion.norrman@risoe.dk |
| |
Abstract: | ![]() Silicon grafted monodisperse poly(ethylene glycol) (PEG) silanes with various PEG chain lengths and mixtures of these were systematically analyzed with static time-of-flight secondary ion mass spectrometry (TOF-SIMS). The mass spectra show differences in the various relative signal intensities, an observation that was used to elucidate important aspects of the grafting process. The relationship between PEG-silane fragment ion abundances and Si(+) ion abundances were used to (i) qualitatively describe layer thicknesses of grafted mixtures of PEG-silanes on silicon, (ii) construct a calibration curve from which PEG chain length (or molecular mass) can be determined and (iii) quantitatively determine surface mixture compositions of grafted monodisperse PEG-silanes of different chain lengths (3, 7 and 11 PEG units). The results suggest that discrimination does take place in the adsorption process. The PEG-silane with the shorter PEG chain is discriminated for mixtures containing PEG3-silane, whereas the PEG-silane with the longer PEG chain is discriminated in PEG7/PEG11-silane mixtures. The origin of this difference in adsorption behavior is not well understood. Aspects of the grafting process and the TOF-SIMS analyses are discussed. |
| |
Keywords: | time‐of‐flight secondary ion mass spectrometry quantification poly(ethylene glycol) silanes protein adsorption |