Flow-injection determination of L-histidine with an immobilized histidine oxidase from Brevibacillus borstelensis KAIT-B-022 and chemiluminescence detection.
Authors:Nobutoshi Kiba  Ayumi Koga  Masaki Tachibana  Kazue Tani  Hitoshi Koizumi  Tetsuo Koyama  Akira Yamamura  Kunio Matsumoto  Tohru Okuda  Koki Yokotsuka
Affiliation:Department of Applied Chemistry, Faculty of Engineering, University of Yamanashi, Takeda 4-3-11, Kofu 400-8511, Japan.
Abstract:
A chemiluminometric flow injection analytical system for the quantitation of L-histidine is described. Histidine oxidase (EC 1.4.3.-) from Brevibacillus borstelensis KAIT-B-022 was immobilized on tresylated poly(vinyl alcohol) beads and packed into a stainless-steel column. The hydrogen peroxide produced was detected chemiluminometrically by a flowthrough sensor containing immobilized peroxidase (EC 1.1 1.1.7). The maximum sample throughput was 10 h(-1). The calibration graph was linear from 0.05 to 5 mM; the detection limit (signal to noise ratio = 3) was 0.01 mM. The activity of immobilized histidine oxidase reduced to 65% of the initial value after 350 injections. The system was applied to the determination of L-histidine in fish meat, such as salmon, tunny, bonito, and mackerel.