Variable temperature system using vortex tube cooling and fiber optic temperature measurement for low temperature magic angle spinning NMR |
| |
Authors: | Martin Rachel W Zilm Kurt W |
| |
Affiliation: | Department of Chemistry, Yale University, P.O. Box 208107, New Haven, CT 06520-8107, USA. |
| |
Abstract: | We describe the construction and operation of a variable temperature (VT) system for a high field fast magic angle spinning (MAS) probe. The probe is used in NMR investigations of biological macromolecules, where stable setting and continuous measurement of the temperature over periods of several days are required in order to prevent sample overheating and degradation. The VT system described is used at and below room temperature. A vortex tube is used to provide cooling in the temperature range of -20 to 20 degrees C, while a liquid nitrogen-cooled heat exchanger is used below -20 degrees C. Using this arrangement, the lowest temperature that is practically achievable is -140 degrees C. Measurement of the air temperature near the spinning rotor is accomplished using a fiber optic thermometer that utilizes the temperature dependence of the absorption edge of GaAs. The absorption edge of GaAs also has a magnetic field dependence that we have measured and corrected for. This dependence was calibrated at several field strengths using the well-known temperature dependence of the (1)H chemical shift difference of the protons in methanol. |
| |
Keywords: | MAS NMR Sample heating Temperature measurement Temperature sensor magnetic field dependence Variable temperature NMR Vortex tube |