首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Electroanalytical performance of a Bismuth/Antimony composite glassy carbon electrode in detecting lead and cadmium
Authors:Grace W Muna  Emily Barrera  Lucas Robinson  Hiba Majeed  Keon Jones  Abigail Damschroder  Andrew Vila
Institution:Department of Chemistry and Biochemistry, Indiana University, South Bend, 46634 IN
Abstract:This work reports on the electroanalytical performance of a glassy carbon electrode (GCE) modified with antimony and bismuth (Sb/Bi-GCE) in detecting heavy metal ions using lead and cadmium as model analytes. The electroanalytical performance of the Sb/Bi-GCE surface was compared to the bismuth modified glassy carbon electrode (Bi-GCE) as well as the antimony modified glassy carbon electrode (Sb-GCE). The Sb/Bi-GCE exhibited excellent figures of merit compared to Bi-GCE and Sb-GCE surfaces. For example, the limit of detection for lead was 0.01 ppb using Sb/Bi-GCE and 0.1 and 1 ppb on Bi-GCE and Sb-GCE, respectively.
Keywords:antimony  bismuth  differential pulse anodic stripping voltammetry  electrodeposition  scanning electron microscopy